The LogTag ® TIC20 is a disposable electronic temperature indicator for use in monitoring temperature & time statistics in pharmaceutical transport applications of up to 40 days duration.
 
The LogTag ® TIC20 is factory pre-configured with one of two monitoring profiles to meet the WHO PQS parameters.
• Clear indication of temperature history/alarm status at all times
• Up to 40 days temperature records can be viewed on the display
• Records temperature from -30C to +60C
• Meets the WHO PQS performance specification EO6/TR07.2.
• Review button to display daily statistics in sequence.
 
Power Source3V Li-Mg Battery
Battery LifeMinimum storage life of 18 months before ‘start'
Monitoring period: 20 days
Minimum accessibility (display) period of 6 months after ‘stopped'
Real Time ClockBuilt-in real time clock
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day)
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C)
Size73mm(H) x 54.5mm(W) x 8.6mm(T)
Weight30g
Case MaterialPolycarbonate
Sensor Measurement Range-30°C to +60°C (-22°F to +140°F)
Operating Temperature Range-30°C to +60°C (-22°F to +140°F)
Storage Temperature Range-30°C to +65°C (-22°F to +149°F)
Rated Temperature Reading AccuracyBetter than ±0.5°C for -10°C to +40°C
Better than ±1.0°C for -30°C to -10°C
Better than ±1.0°C for +40°C to +60°C
Rated Temperature Reading ResolutionLess than 0.1°C for -30°C to +60°C
Less than 0.2°C for -30°C to +40°C
1.0°C from -40°C to +60°C
Sensor Reaction TimeTypically less than 7 minutes (T90) in moving air (1m/s)
Recording CapacityMinimum & maximum temperature for each of the 20 days
1st activation of each alarm for each day including trigger time and duration of each excursion
Please note the TICT does not store each individual temperature reading taken
Sampling Interval1-15 minutes, factory set to 6 minute interval for profile 0006 (i.e. 0.1hour)
Logging Start OptionsPush button start or specific date & time
Start DelayStart delay of 1-99 minutes, factory set to 60 minutes for profile 0006
Recording IndicationAlarms are displayed on a custom LCD
EnvironmentalIP64
  • Thumbnail of document Data Sheet - TIC20 Temperature Indicator With 20 Day Display
    Download

    Data Sheet - TIC20 Temperature Indicator With 20 Day Display

    Download
  • Thumbnail of document Guide - TIC20 Temperature Indicator With 20 Day Display
    Download

    Guide - TIC20 Temperature Indicator With 20 Day Display

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more