Skip to main content
Press Enter
Kaye

Validator AVS

Benchmark system w/ patented asset centric data management, features a new flexible approach to validation

Have questions? We're here to help!
Have questions? We're here to help!

Advanced Validation Technology

The Kaye Validator AVS (Advanced Validation System) combines accurate sensor measurements with all GMP requirements for calibration and traceability to national standards, while generating compliant reports and managing the validated assets and validation equipment. The Validator AVS offers improved hardware features, an easy to use data handling concept and an intuitive, user-friendly software interface.

Reducing Your Cost of Ownership


With our integrated validation system and improved data management concept, the Kaye Validator AVS saves time and money by simplifying the entire validation process.

Lifting Validation to the Next Level

The Kaye Validator AVS introduces a completely new validation concept. The AVS console will be the common platform for all validation products. It is dedicated for validation tasks only – so no worries about operating systems. The Software is preloaded – with a customized interface for validation equipment, simplifying the validation process.

Autoclave Temperature Mapping 
 
Maintaining product quality and regulatory requirements is critical, and with Kaye's autoclave temperature mapping solution, you can seamlessly achieve uniform and accurate temperature distribution throughout your autoclave. By strategically placing Kaye's AVS Loggers within the autoclave chamber, precise temperature data is captured at multiple locations by our autoclave temperature mapping process. Visualize the autoclave temperature mapping using Kaye's AVS software, by providing a graphic of the autoclave and then highlighting the locations of the AVS Loggers. With Kaye's solution, you can confidently demonstrate compliance, enhance process control, and ensure consistent product quality. 
 
Common Reporting Tool Software
 
Sometimes the reporting and post analysis of your validation studies are better done in the convenience of your office on a separate PC. That is possible with the Kaye Common Reporting Tool that opens study files generated with Kaye ValProbe RT and Kaye Validator AVS systems. This provides added flexibility and efficiency with the use of your Validation Console. The installation requires Windows 10 with English (US) settings and keyboard.

 

 

Validator AVS

• Enhanced data handling and redundancy (4-levels)
• Expanded input capacity, types, and range
• Improved scan time 2/sec for 48 inputs (EN554/ISO17665)
• Stand-alone operation with Console docking station

Validator AVS Console

• Hardened, rugged Console pre-loaded with all Kaye software
• Easy-to-operate / state-of-the art intuitive user interface
• Ethernet/ WiFi/ Docking Station connections
• Asset centric concept for process equipment and Kaye hardware
• Console capable of running existing Kaye product software
• Improved Reporting Analysis Tool –backward compatible
• Simplified validation (GAMP-5)

Analog InputUp to 48
ThermocouplesType T, J, K,E,B,R,N,S: 0.1°C; T+ limited range 0.01°C resolution
Scanning Speed36 channel in 1 sec and 48 channels in 2 Sec
Internal Memory4 gb for data collection
Input Impedance10KΩ. Source greater than 10KΩ produces open circuit indication
Common Mode Rejection160 db (8 inputs/sec) @ line frequency
145 db (12 inputs/sec) @ line frequency
140 db @ DC
Max. Common Mode Voltage100V pk ch-to-ch350V pk ch-to ch to frame ground
Normal Mode Rejection82 db @ 60 Hz (8 inputs/sec)69 db @ 60 Hz (12 inputs/sec)
Voltage Input0 to 10 VDC
Resolution1:72,000
Voltage Input Accuracy 30 days: ±(0.003% of reading + 2 counts + 4 microvolts)
1 year: ±(0.006% of reading + 2 counts + 4 microvolts)
Sensitivity0.5 microvolts/count on most sensitive range
Voltage Temp. Coef.±(0.1 microvolts + 0.001% reading)/°C
Compensator Temp. Coef.±0.01°C per °C
Input Terminal Temperature Non-uniformity±0.1°C from calibrated terminal
Input Ranges-6 to 30mV, -12 to 60mV, -60 to 300mV, -2 to 10V
EnvironmentalTemperature: 0 to 50°C (32 to 122°F)
Relative humidity: 95% non-condensing
Power90 to 250 VAC, 50/60 Hz
Fuse Rating4A Slow Blow
Size190H X 411W X 381 mm D (457 mm with SIM)
7.5 in H x 16.2 in W x 15 in D (18 in with SIM)
Weight10.60 kg (23.4 lbs)
BatteryLithium ion with minimum 3 hours of battery backup

• Steam Sterilizers (Autoclaves)  
• Dry Heat Sterilizers
• Steam in Place (SIP)
• Water Cascade/Fall Sterilizers
• Incubators
• Stability Chambers
• Freezers
• Freeze Dryer/Lyophilisation
• Vessels

  • Thumbnail of document Brochure - Validator AVS
    Download

    Brochure - Validator AVS

    Download
  • Thumbnail of document Manual - Validator AVS
    Download

    Manual - Validator AVS

    Download

Request a Service Quote

Get a customized quote for calibration, validation, repair, or training services designed to keep your instruments accurate and compliant.
Learn more

Browse By Brands

Uncover the brands that set the standard for quality and dependability in every industry we serve.
Learn more